Home Precision Environment Multi-Channel Airflow Measurement High-Temperature Multi-Channel Anemomaster PRO / PRO PLUS 1595–1596
High-Temperature Multi-Channel Anemomaster PRO / PRO PLUS 1595–1596 product image thumbnail
High-Temperature Multi-Channel Anemomaster PRO / PRO PLUS 1595–1596 workflow diagram thumbnail
High-Temperature Multi-Channel Anemomaster PRO / PRO PLUS 1595–1596 application overview thumbnail
High-Temperature Multi-Channel Anemomaster PRO / PRO PLUS 1595–1596 application overview thumbnail
High-Temperature Multi-Channel Anemomaster PRO / PRO PLUS 1595–1596 application overview thumbnail
High-Temperature Multi-Channel Anemomaster PRO / PRO PLUS 1595–1596 application overview thumbnail
High-Temperature Multi-Channel Anemomaster PRO / PRO PLUS 1595–1596 application overview thumbnail
KanomaxHigh-Temperature Multi-Channel Anemometer

High-Temperature Multi-Channel Anemomaster PRO / PRO PLUS 1595–1596

Expandable touchscreen platform for simultaneous velocity and temperature measurement at multiple points in furnaces, ovens and thermal test chambers.

Temperature
Up to 500°C
Measurement
Multi-point velocity + temperature
Expansion
Model/configuration dependent
Outputs
Digital plus optional analog units
High-temperature probesMulti-channel architectureTouchscreen operationWindows data software
ISO 13579 method contextNFPA 86 method context

Products

ProductProduct lineApplication areaSelection note
1595 / 1596High-temperature multi-channel systemConfigurable simultaneous velocity and temperature measurementOvens, furnaces, drying systems and thermal-process development

Key Features

High-temperature operation

Compatible probe configurations measure air velocity and temperature at temperatures up to 500°C.

Scalable channel architecture

Build a system around the number of measurement points and probe types required by the application.

Mixed probe configurations

Combine ambient and high-temperature probes where a process includes multiple thermal zones.

Software and output expansion

Use Windows software for live monitoring and optional analog units for voltage or current outputs.

Technical Specifications

ParameterUnitValue / Description
SYSTEM
Measured parametersAir velocity and temperature
Maximum process temperature°C500, probe dependent
User interfaceTouchscreen main unit
PC softwareReal-time monitoring, logging and text/Excel-compatible export
OPTIONAL ANALOG UNIT
Channels per unit12
Voltage outputV0–5
Current outputmA4–20
System channel countDepends on model and configured expansion; confirm final architecture

Values shown are manufacturer specifications or documented configuration details. Confirm final model, probe, accessories, regional availability and calibration scope before ordering.

Applications

Multi-point circulation mapping

Compare velocity and temperature across the working volume.

Thermal-zone verification

Identify non-uniform airflow or recirculation at multiple locations.

Process optimization

Correlate multi-point airflow with temperature and product quality.

Qualification and development

Document spatial conditions during equipment testing.

Similar Products