CPA 341 Cleanroom Particle Analyzer
Portable or stationary single-particle light-scattering counter for low-concentration aerosols in gaseous samples — covering 0.3 to 10 µm at a fixed sampling flow of 28.3 L/min for cleanroom qualification, acceptance testing and filter measurements.
Model Overview
| Model | Measurement principle | Core specification | Core fit |
|---|---|---|---|
| CPA 341 | Single-particle 90° light-scattering counter with internal flow generation | 0.3–10 µm, five fixed size classes and 28.3 L/min sample flow | Cleanroom qualification, acceptance measurements, filter-integrity support and fractional-efficiency testing |
Key Features
Stable 28.3 L/min sampling
An internal side-channel blower maintains a 28.3 L/min (1 cfm) sampling flow, supporting both mobile measurements and stationary monitoring of low-concentration aerosols.
Five fixed particle-size classes
The optical counter measures particles from 0.3 to 10 µm in five preset size classes, providing particle-number and size information for cleanroom and filter-test workflows.
Programmable cleanroom workflows
Up to 999 measurement programs can store measurement and pause times, measurement positions, cycle counts and cleanroom-class alarm settings for repeatable qualification routines.
Touchscreen, alarms and data handling
The glove-operable touchscreen supports single, repeat and continuous measurement modes. Results can be displayed and printed, transferred through Ethernet and processed with CRQWin software.
Technical Specifications
| Parameter | Unit | Value / Description |
|---|---|---|
| MEASUREMENT SYSTEM | ||
| Measurement principle | — | Single-particle light scattering with 90° collection optics |
| Particle-size range | µm | 0.3 – 10 |
| Number of size classes | — | 5, fixed |
| Sample flow | L/min | 28.3 (1 cfm) |
| Flow generation | — | Internal side-channel blower |
| Zero-count rate | counts | < 1 count per 5 minutes, evaluated according to ISO 21501-4 |
| Counting efficiency | % | 50% ± 20% at 0.3 µm, evaluated according to ISO 21501-4 |
| OPERATION AND DATA | ||
| Display and control | — | Touchscreen |
| Stored measurement programs | — | Up to 999 |
| Measurement modes | — | Single, repeat and continuous measurement |
| Measurement / pause time | — | 1 second to 99 hours |
| Alarm | — | Configurable for each cleanroom class |
| Interface | — | Ethernet |
| Result handling | — | Tabular display, thermal-printer output and CRQWin processing |
| Inlet connection | mm | Ø 8 |
| PHYSICAL AND ELECTRICAL | ||
| Power supply | — | 100 – 240 V AC, 50/60 Hz |
| Dimensions (W × H × D) | mm | 225 × 255 × 220 |
| Weight | kg | 4 |
Specifications are based on the manufacturer data sheet dated March 2026. Confirm the final delivery scope, accessories and calibration documentation before ordering.
Applications
Classification and validation
Particle-concentration measurements for cleanroom qualification and validation according to ISO 14644-1/-3, GMP Annex 1 and VDI 2083.
Cleanrooms and workbenches
Mobile acceptance measurements of cleanrooms, clean-air workbenches and ventilation systems with high cleanliness requirements.
Integrity and fractional efficiency
Upstream concentration measurement and downstream leak scanning in filter-integrity workflows, plus determination of fractional separation efficiency for filters and separators.