Test flow
Up to 1,200 m³/h
Scanning
Manual probe scan
Specimen size
Up to 1,220 × 610 × 300 mm
Aerosols
DEHS / PSL
Manual scanning Automated measurement Compact system HEPA / ULPA EN 1822
ISO 29463-3 ISO 29463-4 ISO 29463-5 EN 1822-3 EN 1822-4 EN 1822-5

Model Overview

Model System role Primary measurements Standards / application fit
AFS 152 Manual HEPA/ULPA scanning system Manual probe traversal with automated particle measurement and report generation Compact production control and classification of HEPA/ULPA filters

Key Features

Manual scan, automated measurement

The operator moves the probe while the system records and evaluates particle measurements.

Scanned-area documentation

Test software documents the scanned filter area for traceable quality records.

Standard-compliant reports

Reports are structured for ISO 29463 and EN 1822 production-control requirements.

Compact footprint

The system is designed for laboratories and production areas with limited space.

HEPA/ULPA classification

Supports leak scanning, integral efficiency and differential-pressure measurements.

Upgrade capability

Suitable systems can be upgraded with the automated AFS 152/A traverse.

Technical Specifications

Parameter Unit Value / Description
TEST PERFORMANCE
Test flow rate m³/h Maximum 1,200
Differential-pressure range Pa Maximum 400
Maximum specimen size mm 1,220 × 610 × 300
Scanning Manual scan with automated measurement
SYSTEM
Test aerosols DEHS and PSL
Compressed-air supply bar 6
Power supply 230 VAC, 10 A
Outputs Scanned-area documentation and standard-compliant report

Values shown are based on the standard system configuration. Confirm the final test method, specimen fixtures, instrumentation, utilities and calibration scope before order placement.

Applications

HEPA/ULPA production control

Manual scanning with automated data acquisition and evaluation.

Filter classification

Testing according to ISO 29463 and EN 1822.

Leak detection

Systematic manual scan of filter media and frame for local defects.

Compact filter testing

Space-efficient testing for panel-filter development and acceptance.